Tri-Parameter Portable Patient Monitor
Features:
●Basic parameters (ECG, NIBP, SpO2, TEMP, RESP) to reach general clinical needs
●Bright 8" TFT, easy to view
●Compact and light for transfer / transport
●Pitch tone to provide audibly recognized SpO2
●Up to 72-hour trend to review previous patient data
Standard configuration of H5: ECG, SpO2, NIBP, 2-TEMP, RESP
Option: Thermal Recorder, ETCO2, 2IBP
Technical Specification
- Safety:IEC60601(GB9706)
- Power Supply:100-240VAC,47-63Hz
- Patient Type:Adult,Pediatric and Neonate
Display
- 8.4〞Color TFT display
- Indicator: Alarm indicator,Power indicator,Battery indicator,Patient type indicator,Mute indicator,NIBP measuring indicator,Pulse bar indicator
NIBP
- Technique: Automatic Oscillometry
- Range: Adult: 10-270mmHg
Child: 10-235mmHg
- Neonate: 10-135mmHg
- Accuracy: Static ± 3mmHg
- Clinic: Average error: ±5mmHg
- Standard Deviation: ≦8 mmHg
- Unit: mmHg. kPa
- Pulse rate range: 40-240 bpm
SpO2
- Measurement Range: (0~100)%
- Resolution: 1%
- Accuracy: ±2% (70%~100%)
±3% (35%~69%)
- Unspecified: (0~34%)
ECG
- Lead type: 5-lead/3-lead
- Input: RA, LA, RL, LL, V
- Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
- Accuracy: ±1% bpm or ±1 % (whichever is greater)
- Protection: withstand 4000VAC/50Hz voltage in isolation against electrosurgical and defibrillation
- S-T detection: YES, Arrhythmia analysis: YES
- Alarm: YES, audible and visual alarm, alarm events recallable
RESP
- Method:RA-LL impedance
- Measurement range:Adult:0~120rpm
Pediatric/Neonatal:0~150rpm
- Apnea alarm:YES
Pulse Rate
- Measurement Range: 20~250bpm
- Resolution: 1 bpm
- Accuracy: ±1% or ± 1 bpm, whichever is the greater
TEMP
- Range:0-50℃
- Unit:Celsius,Fahrenheit
- Accuracy:±0.1℃(25℃-50℃)
Microstream EtCO2
- Measurement way: Infrared spectrum
- Measurement mode: Side stream
- Sample Flow Rate: 50ml/min ±10ml/min
- Measurement range: (0~19.7)%, (0~150mmHg)
- Resolution: 0.1% or 1 mmHg
- Unit: %, mmHg, kPa
- Accuracy: ±2 mmHg 0 mmHg to 40 mmHg,
±5% of reading 41 mmHg to 70 mmHg,
±8% of reading 71 mmHg to 100 mmHg,
±10% of reading 101 mmHg to 150 mmHg,
(760 mmHg, temperature is 25℃)
(When RR>80 rpm, all the range is ± 12% of reading)
Alarm
- Lever:3-level,low,medium and high
- Indication:sound and light
- Setup:Default and Custom
- Silence:all alarms can be silenced
Review
- 480 hours trend data
- ≥2400 groups of NIBP measuring data review
Environment
- Temperature:
Operation:0℃-50℃
Storage/Transportation:-40℃-70℃
- Humidity:
Operation:10℃-95℃,noncondensing
Storage/Transportation:0℃-93℃,noncondensing
- Atmospheric Pressure:
Operation:700-1060hPa
Storage/Transportation:500-1060hPa
Size and weight
Packing size: 42x32x33cm
Net weight: 4kg
Gross weight: 5kg