Portable Patient Monitor


Features:

Pitch tone to provide audibly recongnized SpO2
Most Popular Basic 5 parameters(ECG, RESP, NIBP, SpO2, TEMP) to reach general clinical needs
12.1" TFT display, easy to view, Full Touch Screen as option
Arrhythmia and ST analysis to provide rapid cardiac diagnosis
Up to 480-hour trend to review previous patient data
Drug calculation and titration, useful clinical tools
NO Fan design suitable for night time operation and saving energy


Technical Specification


Standard configuration of H8: ECG,HR,RESP,SpO2, NIBP, 1-TEMP, Lithium Battery

Optional parameter& configuration of H8: 2-TEMP, Thermal recorder, SideStream EtCO2, MainStream EtCO2, Full Touch Screen,IBP


Size and Weight
- Product size: 31.8x26.4x15.2cm
- Packing size: 38x38x25cm
- Net weight: 4.5kg
- Gross weight: 6.5kg

Power supply
- Power Voltage: AC 100-240V 50/60Hz
- Power Input: ≦85VA
- Safety class: Category I

Display
- 12.1〞Color TFT display
- Resolution: 800x600 pixels or higher

Battery
- Type: Rechargeable Lithium battery
- Operating time under the normal use and full
charge: ≧120minutes

Thermal Recorder
- Method: Thermal dot array
- Paper width: 50mm(1.97 in)
- Paper Speed: 12.5/25/50(single track or tri-track
mm/sec)
- Traces: Single track or Tri-track

Alarm
- Three Levels: Low, medium and high
- Indication: Auditory and visual
- Setup: Default and custom
- Silence: All alarm can be silenced
- Volume: 45-85 dB measured at 1 meter

Storage & Reviewing
- Trend: 480-hour tabular and graphic trends & data
storage
- NIBP: ≥2400 groups NIBP measurement data
ECG
- Lead mode: 5-lead or 3-lead able
- Leda ion:Ⅰ;Ⅱ;Ⅲ;avR;avL;avF;Vx;CAL
- Sweep speed: 12.5mm/s,25mm/s,50mm/s
- Accuracy: ±1bpm or ±1%,whichever is greater
- Heart beat range: Adult:15~300bpm
Pediatric/Neonatal:15~350 bpm

HR
- Range: Adult 10~300bpm
Pediatric & Neonate: 10~350bpm
- Refreshing time:  ≤50 bpm Per2 pulses
50~120bpm per 4 pulses
≧120bpm per 6 pulses
- Resolution:1bpm
- Accuracy: ±1% or ±1 bpm,whichever is greater

ST segment
- Measurement range: -2.0mV~2.0mV
- Resolution:0.01mV

RESP
- Method: RA-LL impedance
- Measurement range: Adult: 0-120rpm
Pediatric/neonatal: 0-150rpm
- Apnea alarm: YES

NIBP
- Method: Automatic oscillometry
- Range of measurement:
Adult: 10~270mmHg
Child: 10~235mmHg
Neonate: 10~135mmHg
- Accuracy: Static:±2% or ±3mmHg,whichecer is greater
- Unit: mmHg, kPa
- Over-pressure protection: YES
Standard SpO2 ( Digital )
- Measurement Range:0-100%
- Accuracy: At70-100%,±2%
At0-69%,unspecified

PR
- Measurement Range: 25-300bpm
- Accuracy: ±1% or ±1bpm,whichever is greater

TEMP
- Measurement Range: 25.0℃~50.0℃
- Accuracy: ±0.2℃ (exclusive of probe)
- Unit: Celsius(℃),Fahrenheit(℉)
- Connecting cable: Compatible with YSI-400 serial

IBP
- Channel : 2
- Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
- Measuring and alarm range: -50~350mmHg
- Press Sensor
Sensitivity: 5 uVmmHg          Impedance: 300-3000Ω
- Resolution : 1mmHg
- Accuracy: ±2% or 1mmHg which great
- Actualization Interval: about 1 Sec

Environment
- Operating temperature:0~+40℃
- Transportation and Storage temperature:-20℃ to +50℃
- Relative humisity:Working ≦85%
Transportation and storage ≦93%
- Atmospheric pressure:860hPa to 1060hPa
Transportation and storage 500-1060 hPa
- Analyze function
- Lan or wireless function
- 7 waveforms on the screen