Welcome to Sino-Hero(Shenzhen)Bio-Medical Electronics Co.,Ltd
S60 Vista
Catalog Download
Product Introduction

Vista Vital Sign Patient Monitor


Features:

7" TFT high-resolution color display with sharp view
Parameters Highlight: SpO2 (Digital, Nellcor, and Masimo able), NIBP, Quick Temp, EtCO2 (Microstream)
Super compact and lightweight design for instant availability
Exceptional data storage capacity both for continuous monitoring and spot check
Built-in Li-ion battery life last up to 8 hours
Wire net working with Central Monitoring Systems


Technical Specification

Standard Configuration: NIBP, Digital SpO2, PR, Lithium Battery
Option: Nellcor SpO2, Masimo SpO2, Quick TEMP, Micro Stream EtCO2, Thermal recorder, Full Touch Screen, BarCode scanner.

Size and Weight
- Product size: 16x13x26cm
- Packing size: 38x33x28cm
- Net weight: ≦2.5kg
- Gross weight: 4.5kg

Display
- Type: 7〞 Color Anti-glare TFT-LCD
- Resolution: 234×480 pixels

Power supply
- Input voltage:  AC: 100-240V,  50Hz/60Hz
- Input power: 70VA
- Fuse: T1.6AL/250V, 2φ5x20mm

Battery
- Type:  11.1V 4000mAh rechargeable lithium battery
- Operating time: >8 hours
- Charge time: 6h to 100% (Standby)

Data Storage
- Clinic Mode:
Patient quantity: 1000
Recording number: 16000
Single patient clinic record number: 1-16000
- Monitor Mode:
Patient quantity: 1
Trend data: resolution 1Min: can store 96 hours
resolution 5Min: can store 480 hours
resolution 10 Min: can store 960 hours
Alarm events: 1000
NIBP measurement record: 5000

Environmental Specifications
- Operating
Temperature: (5~40) ℃
Humidity: ≦85% (noncondensing)
Atmospheric Pressure: (860~1060) hPa
- Transportation and Storage
Temperature: (-20 ~ +55)
- Humidity: 93% (noncondensing)
- Atmospheric Pressure: (500~1060) hPa

PR
- Measurement Range: 20~250bpm
- Resolution: 1 bpm
- Accuracy:  ±1% or ± 1 bpm, whichever is the greater

Digital SpO2
- Measurement Range:  (0~100)%
- Resolution: 1%
- Accuracy:  ±2% (70%~100%)
±3% (35%~69%)
- Unspecified:  (0~34%)

Nellcor-SpO2 (Option)
- Measurement Range:  (0-100)%
- Resolution: 1%
- Accuracy:  ±2% (70%~100%) (adult)
±3% (70%~100%) (neonate)
±2% (70%~100%) (low perfusion)
- Unspecified:  (0~69%)

Quick TEMP
- Ambient temperature: 10℃-40℃
- Measurement range: 34.0℃-43.0℃
- Measurement part: Ear cavity
- Unit:℃,℉
- Resolution: 0.1 ℃/℉
- Accuracy: Between 35.0℃-42.0℃: ≦ ±0.2℃
Beyond 35.0℃-42.0℃: ≦ ±0.3℃
- Clinical repeat accuracy: ≦ ±0.3℃
- Measurement time: ≦ 0.8s
- Measurement interval time: 12s
- Standby time:40s
- Memory storage: 10

NIBP
- Technique: Automatic Oscillometry
- Range: Adult: 10-270mmHg
Child: 10-235mmHg
- Neonate: 10-135mmHg
- Accuracy: Static ± 3mmHg
- Clinic: Average error: ±5mmHg
- Standard Deviation: ≦8 mmHg
- Unit: mmHg. kPa
- Pulse rate range: 40-240 bpm

Microstream EtCO2
- Measurement way:   Infrared spectrum
- Measurement mode: Side stream
- Sample Flow Rate: 50ml/min ±10ml/min
- Measurement range: (0~19.7)%, (0~150mmHg)
- Resolution: 0.1% or 1 mmHg
- Unit: %, mmHg, kPa
- Accuracy: ±2 mmHg  0 mmHg to 40 mmHg,
±5% of reading 41 mmHg to 70 mmHg,
±8% of reading 71 mmHg to 100 mmHg,
±10% of reading 101 mmHg to 150 mmHg,
(760 mmHg, temperature is 25℃)
(When RR>80 rpm, all the range is ± 12% of reading)