Vista Vital Sign Patient Monitor
Features:
●7" TFT high-resolution color display with sharp view
●Parameters Highlight: SpO2 (Digital, Nellcor, and Masimo able), NIBP, Quick Temp, EtCO2 (Microstream)
●Super compact and lightweight design for instant availability
●Exceptional data storage capacity both for continuous monitoring and spot check
●Built-in Li-ion battery life last up to 8 hours
●Wire net working with Central Monitoring Systems
Technical Specification
Standard Configuration: NIBP, Digital SpO2, PR, Lithium Battery
Option: Nellcor SpO2, Masimo SpO2, Quick TEMP, Micro Stream EtCO2, Thermal recorder, Full Touch Screen, BarCode scanner.
Size and Weight
- Product size: 16x13x26cm
- Packing size: 38x33x28cm
- Net weight: ≦2.5kg
- Gross weight: 4.5kg
Display
- Type: 7〞 Color Anti-glare TFT-LCD
- Resolution: 234×480 pixels
Power supply
- Input voltage: AC: 100-240V, 50Hz/60Hz
- Input power: 70VA
- Fuse: T1.6AL/250V, 2φ5x20mm
Battery
- Type: 11.1V 4000mAh rechargeable lithium battery
- Operating time: >8 hours
- Charge time: 6h to 100% (Standby)
Data Storage
- Clinic Mode:
Patient quantity: 1000
Recording number: 16000
Single patient clinic record number: 1-16000
- Monitor Mode:
Patient quantity: 1
Trend data: resolution 1Min: can store 96 hours
resolution 5Min: can store 480 hours
resolution 10 Min: can store 960 hours
Alarm events: 1000
NIBP measurement record: 5000
Environmental Specifications
- Operating
Temperature: (5~40) ℃
Humidity: ≦85% (noncondensing)
Atmospheric Pressure: (860~1060) hPa
- Transportation and Storage
Temperature: (-20 ~ +55)
- Humidity: 93% (noncondensing)
- Atmospheric Pressure: (500~1060) hPa
PR
- Measurement Range: 20~250bpm
- Resolution: 1 bpm
- Accuracy: ±1% or ± 1 bpm, whichever is the greater
Digital SpO2
- Measurement Range: (0~100)%
- Resolution: 1%
- Accuracy: ±2% (70%~100%)
±3% (35%~69%)
- Unspecified: (0~34%)
Nellcor-SpO2 (Option)
- Measurement Range: (0-100)%
- Resolution: 1%
- Accuracy: ±2% (70%~100%) (adult)
±3% (70%~100%) (neonate)
±2% (70%~100%) (low perfusion)
- Unspecified: (0~69%)
Quick TEMP
- Ambient temperature: 10℃-40℃
- Measurement range: 34.0℃-43.0℃
- Measurement part: Ear cavity
- Unit:℃,℉
- Resolution: 0.1 ℃/℉
- Accuracy: Between 35.0℃-42.0℃: ≦ ±0.2℃
Beyond 35.0℃-42.0℃: ≦ ±0.3℃
- Clinical repeat accuracy: ≦ ±0.3℃
- Measurement time: ≦ 0.8s
- Measurement interval time: 12s
- Standby time:40s
- Memory storage: 10
NIBP
- Technique: Automatic Oscillometry
- Range: Adult: 10-270mmHg
Child: 10-235mmHg
- Neonate: 10-135mmHg
- Accuracy: Static ± 3mmHg
- Clinic: Average error: ±5mmHg
- Standard Deviation: ≦8 mmHg
- Unit: mmHg. kPa
- Pulse rate range: 40-240 bpm
Microstream EtCO2
- Measurement way: Infrared spectrum
- Measurement mode: Side stream
- Sample Flow Rate: 50ml/min ±10ml/min
- Measurement range: (0~19.7)%, (0~150mmHg)
- Resolution: 0.1% or 1 mmHg
- Unit: %, mmHg, kPa
- Accuracy: ±2 mmHg 0 mmHg to 40 mmHg,
±5% of reading 41 mmHg to 70 mmHg,
±8% of reading 71 mmHg to 100 mmHg,
±10% of reading 101 mmHg to 150 mmHg,
(760 mmHg, temperature is 25℃)
(When RR>80 rpm, all the range is ± 12% of reading)