Multi-Parameter Patient Monitor


Features:

12.1 " color TFT display with maximum 10 waveforms
Power software function include OxyCRG, Large Font, Short Trend, and Drug calculations
Pitch tone to provide audibly recognized SpO2
Maximum 168-hour trends to review all parameters\\\\' information
Rechargeable battery for long time use
Wire net working with Central Monitoring Systems


Technical Specification


Size and Weight
- Product size: 31.8x26.4x15.2cm
- Packing size: 43x32x35cm
- Net weight: 4.5kg
- Gross weight: 6.5kg

Power supply
- Power Voltage: AC 100-240V 50/60Hz
- Power Input: ≦85VA
- Safety class: Category I

Display
- 12.1”Color TFT display
- Resolution:800x600 pixels

Battery
- Type: Rechargeable Acid Battery(Option Lithium
battery)
- Operating time under the normal use and full charge:≧
120minutes
- 2 batteries ≧240 minutes (Option Lithium battery will
double operate time)


Thermal Recorder
- Method: Thermal dot array
- Paper width:50mm(1.97 in)
- Paper Speed:12.5/25/50(mm/sec)
- Traces: Maximum 3 tracks

Alarm
- Three Levels: Low, medium and high
- Indication: Auditory and visual
- Setup: Default and custom
- Silence: All alarm can be silenced
- Volume: 45-85 dB measured at 1 meter

Trend
- Trend: 168 hours
- Parameters option: HR, SpO2, NIBP, PR, RESP,
EtCO2, Temp1, Temp2, AA, N2O,O2, IBP1, IBP2, ST.
- Cycle intervals of trend storage1 min, 2 min, 3 min, 4
min, 5 min, 10 min, 15 min, 20 min, 25 min, 30 min.

Storage & Reviewing
- ECG: 30 minutes one important lead’s ECG waveform
- Alarm: 1800 groups alarm events reviewing
- NIBP: 1000 groups NIBP measurement
- Arrhythmia: 128 groups data(8 seconds ECG
waveform)
- Power-off storage: 72 hours trend data & 1 ECG wave
form (Option)

ECG
- Lead Mode:3-leads ECG input
5-leads ECG input
- Lead ion:Ⅰ,Ⅱ,Ⅲ
Ⅰ,Ⅱ,Ⅲ,aVR,aVL,aVF,V-
- Gain: 2.5mm/mV(×0.25), 5mm/mV(×0.5), 10mm/m
V(×1), 20mm/mV(×2), 40mm/mV(×4),Auto
- CMRR: Monitor mode≧105dB
Surgery mode≦105dB
Diagnostic mode≧90dB
- Frequency response(-3d B):
Monitor mode 0.5~40Hz
Surgery mode 1~25Hz
Diagnostic mode 0.05~150Hz
- input impedance: ≧5.0Mohm
- ECG signal range:±10.0mV
- Electrode offset potential: ±500mV
- Patient Leakage Current:<10uA
- Standardizing signal:1mV±5%
- Baseline recovery:<5s after Defibrillation.(Mon or Surg
mode)
- Indication of electrode separation: Every electrode(ex
clusive of RL)
- Protection: Breakdown Voltage 4000AVC 50/60Hz;de
fibrillator proof
- Sweep speed:12.5mm/s,25mm/s,50mm/s

HR
- Range:  Adult 10~300bpm
Pediatric & Neonate 10~350bpm
- Refreshing time:≦50 bpm Per2 pulses
50~120bpm per 4 pulses
≧120bpm per 6 pulses
- Resolution:  1bpm
- Accuracy:   ±1% or ±1bpm,whichever is greater
ST segment
- Measurement range: -2.0mV~2.0mV
- Resolution:0.01mV

RESP
- Method: Impedance variation between RA-LL(R-F)
- Gain: ×1, ×2, ×4
- Bandwidth:0.25Hz to 2.0Hz (-3dB)
- Sweep Speed: 6.25mm/s,12.5mm/s,25mm/s
- Measurement Range:0~150 rpm
- Accruacy:±2rpm

NIBP
- Method: Automatic oscillometry
- Range of measurement:
Adult: 10~270mmHg
Child: 10~235mmHg
Neonate: 10~135mmHg

- Accuracy: Static:±2% or ±3mmHg,whichecer is greater
- Unit: mmHg, kPa
- Intervals for AUTO measurement
time:1,2,3,4,5,10,15,20,30,45,60,90 minutes; 2,4,8 hours
- Pulse rate range:40~240bpm

Standard SpO2(Digital)
- Measurement Range:0~100%
- Accuracy: At70~100%,±2%
At0~69%,unspecified
PR
- Measurement Range:25~250bpm
- Accuracy: ±1% or ±1bpm,whichever is greater

Nellcor-SpO2
- Measurement Range:0~100%
- Resolution:1%
- Accuracy:At70~100%:±2%(Adult)
At70~100%:±3%(Neonate)
At70~100%:±2%(Low Perfusion)
At0~69%,unspecified
PR
- Measurement Range:20~300bpm
- Resolution:1bpm
- Accuracy: 20bpm to 250bpm: ±3bpm
251bpm to 300bpm: unspecified

Masimo-SpO2
- Measurement Range: 0% to 100%
- Resolution:1%
- Accuracy:
70% to 100%:±2%(adult/pediatric, on-motion conditions)
70% to 100%:± 3 %( neonate, on-motion conditions)
70% to 100%:± 3 %( motion conditions)
0% to 69%, unspecified
- Average time:2-4s,4-6s,8s,10s,12s,14s,16s

PR
- Measurement Range:25bpm to 240bpm
- Accuracy: ±3bpm(non-motion conditions)
± 5bpm (motion conditions)
- Resolution: 1bpm

TEMP
- Measurement Range: 25.0℃~50.0℃
- Accuracy:  ±0.1℃ (exclusive of probe)
- Unit: Celsius(℃),Fahrenheit(℉)
- Connecting cable: Compatible with YSI-400 serial

IBP
- Channel: 2
- Measurement way: Directly invasive pressure measure
ment
- Sensitivity of transducer: 5uV/V/mmHg, ±2%
- Impedance of transducer: 300 to 3000Ω
- Measurement Range: -50~+300mmHg
- Resolution:  1mmHg
- Unit:   mmHg, kpa, cmH2O
- Accuracy:
Static: ±1mmHg or ±2%, whichever is grater(exclusive of
transducer)
±4mmHg or ±4%, whichever is grater(inclusion of
transducer)
Dynamic:±4mmHg or 4%, whichever is grater
Transducer sites: Arterial Pressure                    (ART)
Pulmonary Artery Pressure    (PA)
Left Atrium Pressure              (LAP)
Right Atrium Pressure            (RAP)
Central Venous Pressure       (CVP)
Intracranial Pressure              (ICP)

EtCO2 (Sidestream)
- Measure method: Infrared spectrum
- Measure Range: 0.0~13.1 %( 0~99.6mmHg)
- Resolution: 1mmHg
- Unit: %,mmHg,kpa
- Accuracy: 0% to 4.9%, ±0.3 %( ±2.0mmHg)
5.0%to 13.1 %,< ±10%of the reading
- Measurement range of awRR: 2~150rpm
- Calibration: Offset calibration:auto,manual,Gain calibration

EtCO2(Mainstream)
- Measure method: Infrared spectrum
- Warm up time: Capnogram displayed in less than 15
seconds,At an ambient temperature of 25℃,full specifications
within 2 minutes.
- Measure Range: 0.0~19.7 %(0~150mmHg)
- Resolution: 1mmHg
- Rise time(10 l/min):≦60ms
- Unit: %,mmHg,kpa
- CO2 Accuracy: 0-40mmHg, ±2mmHg
41-70mmHg, ±5% of reading
71-100mmHg, ±8% of reading
101-150mmHg, ±10% of reading
(at 760 mmHg,ambient temperature of 25℃)
- awRR measurement range:0~150rpm
- awRR measurement Accuracy: ±1rpm

EtCO2 (Microstream)
- Measure method: Infrared spectrum
- Warm up time: Capnogram displayed in less than 20 seconds,
At an ambient temperature of 25℃,full specifications within 2
minutes.
- Measure Range: 0.0-19.7%(0-150mmHg)
- Resolution: 1mmHg
- Unit: %,mmHg, kpa
- CO2 Accuracy: 0-40mmHg, ±2mmHg
41-70mmHg, ±5% of reading
71-100mmHg, ±8% of reading
101-150mmHg, ±10% of reading
(At 760 mmHg, ambient temperature of 25℃)
(When RR>80 rpm, all the rang is ±12% of reading)
CO2 response time: <3S
- awRR measurement range:2~150rpm
- awRR measurement Accuracy: ±1rpm
- Sample Flow Rate:50ml/min ±10ml/min

Multi Gas
- Measure method: Infrared spectrum
- Fi and ET values: CO2 ,N2O ,O2 ,AG(HAL,ISO,NEF,
SEV,DES)
- Resolution:1%
- Unit:%
- Calibration: Room air calibration automatically when changing
airway Adapter (<5 sec)
- Warm-up time:<10S,full accuracy within 1min
- Measurement and alarm range of AG
- Gas                       Range             Accuracy
CO2                      0-10%             ±(0.3% ABS+4%REL)
N2O                      0-100%           ±(2% ABS+8%REL)
O2                        10-100%          ±(2% ABS+2%REL)
HAL,ISO,ENF       0-5%               ±(0.15% ABS+10%REL)
SEV                      0-8%               ±(0.15% ABS+10%REL)
DES                      0-18%             ±(0.15% ABS+10%REL)
- awRR measurement range:  0~150rpm
- awRR measurement Accuracy: ±1rpm
- Rise time(flowing speed 10I/min)CO2≦90ms
O2≦300ms
N2O≦300ms
Hal,Iso,Enf,Sev,Des≦300ms
- Total system response time :<1seconds

Environment
- Operating temperature:0~+40℃
- Transportation and Storage temperature:-20℃ to +50℃
- Relative humisity:Working≦85%
Transportation and storage ≦93%
- Atmospheric pressure:860hPa to 1060hPa
Transportation and storage 500-1060 hPa

Standard configuration of S90:
ECG,HR,RESP,SpO2,NIBP,1-TEMP,Lithium Battery,PR

Optional parameter& configuration of S90:
Nellcor SpO2,Masimo SpO2,2-TEMP,2-IBP,EtCO2(Sidestream,Microstream,Mainstream)
Multigas/O2,VGA output,Thermal recorder.