Product Introduction
Modular Patient Monitor
Features:
● 15”High-resolution Full Touch Screen
● Modular design for flexibility and freedom to pre-configure and major monitoring parameters
● Friendly interface ensures efficient operations during the surgery
● Optional VGA output for slave display in OR
● High-end Parameters: SunTech NIBP,CO,EtCO2,AG
● Exceptional connectivity by high-speed cable or wireless Networking
Technical Specification
Safety
IEC60601-approved, CE marking to MDD93/42/EEC
Dimension and Weight
Dimension: 404(W)×327(H)×188(D)mm
Weight: 6.3kg
Operation Environment
Power Requirement: AC100-240V, 50/60Hz,140VA,1.4A-0.59A
Temperature 0℃~40℃
Humidity 15%~80% (non-condensing)
Patient range Neonate, pediatric and adult patient
Patient range Neonate, pediatric and adult patient
Display: 15" Color TFT LCD
Rolling and refreshing waveform display
Resolution:1024*768
Multi displays able, including
Standard display
Trend coexist display
Large-font display
OxyCRG dynamic view display
Trace: 7 traces
Sweep speed:12.5mm/s,25mm/s,50mm/s
Indicator: Alarm indicator light
Power indicator light
Battery indicator light
QRS beep and alarm sound
Interface: Parameter cable interface
AC Power input socket
Network interface
External VGA interface for connection
Li-ion Battery: Rechargeable
(one piece) 8 hours for charging
(to 90% battery capacity) 2 hours for continuous working
Trend time: 168 hours
Alarm: User-adjustable High, Medium, and Low limits
3-level audible and visual alarm
Networking: Connected to central station system
Recorder: Thermal array
Waveform: 2 Channel (standard) Real time recording
Paper width: 50mm
Print speed: 25mm/s
ECG
3-Lead or 5-Lead able
Input: 3 leadwire cable:RA;LA;LL or R;L;F
5 leadwire cable:RA;LA;RL;LL;V or R;L;N;F;C
Lead ion: 3-lead: I,II,III
5 lead:I,II,III,AVR,AVL,AVF, V
ECG Waveform 2 channel
Gain ion ×0.25,×0.5,×1,×2,x4,Auto
Sweep speed 12.5mm/s, 25mm/s, 50mm/s
Heart Rate for Neonate 15-350BPM
Heart Rate for Adult/Pediatric 15-300BPM
Accuracy 1bpm or 1%, whichever is greater
Resolution 1bpm
Filter Diagnostic mode:0.05-100Hz
Monitor mode:0.5-40Hz
Surgical mode:1.0-25Hz
Protection Withstand 5000VAC/50Hz Voltage in isolation
Against electrosurgical interference
and defibrillation
Alarm range: Neonate:15-350bpm
Adult/Pediatric:15-300BPM
ST segment detection: Measurement range: -2.0mv to 2.0mv
Alarm range: -2.0mv to 2.0mv
Arrythmia analysis: Yes
Alarm Yes,audible and visual alarm, alarm
event recallable
Pace detection Yes
Drug dose calculation Yes
Standby mode Yes
Respiration
Method Thoracic impedance
Measurement range Neonate/Pediatric:0~150rpm
Adult:0-120rpm
Apnea alarm No alarm,10s,15s,20s, 25s,30s,35s,
40s able
Alarm Yes, audible and visual alarm, alarm
events recallable
NIBP
Method: Oscillometric
Measurement modes: Manual/Automatic/Continuous
Displayed parameters Systolic, diastolic and
MAP(mean arterial pressure)
Auto measure time: Adjustable
Measurement unit mmHg/kPa able
Measurement type: Systolic, Diastolic, Mean
Range of Systolic pressure Adult: 40-270mmHg
Pediatric:40-200mmHg
Neonate:40-135mmHg
Range of Diastolic pressure Adult: 10-210mmHg
Pediatric:10-150mmHg
Neonate:10-100mmHg
Range of Mean pressure Adult: 20-230mmHg
Pediatric:20-165mmH
Neonate:20-110mmHg
Accuracy ±6mmHg
Resolution: 1mmHg
Alarm: Systolic, Diastolic ,Mean
Over-pressure protection: Double safety protection
SpO2
Masimo SpO2
Measurement range: 0%~100%
Resolution: 1%
Accuracy: ±2%(70%-100%); unspecified(0-69%)
Alarm range: 0~100%
Pulse Rate Range:20~254bpm
Resolution:1bpm
Accuracy:±3bpm
Alarm range:0~254bpm
Temperature
Measurement range: 0~50ºC
Resolution: 0.1ºC
Accuracy: ±0.1℃
Channel: T1、T2、TD
IBP
Channels 1 or 2 able
Measurement range -50~400mmHg
Resolution 1mmHg
Alarm: Systolic, Diastolic ,Mean
Over-pressure protection: Double safety protection
ETCO2
Measurement range 0-99mmHg Accuracy 0 to 40mmHg ±2mmHg
41 to 76mmHg ± 5mmHg
77 to 99mmHg ±10mmHg Sample flow rate Minimum: 50ml/m-7.5ml/m Maximum: 50ml/m+15ml/m
Start-up time <1min,once the module starts up, it reaches ISO After 1 min,enter the full accuracy mode Respiration rate 0 to 120rpm
Respiration rate accuracy ±2 rpm
Response time <240msec(10% to 90%)
Cardiac Output
Method: Thermodilution
Measurement rage: CO:0.1-20L./min.
TB: 23-43ºC
TI:0-27ºC
Resolution: CO:0.1/min.
TI:0.1ºC
TB: 0.1ºC
Accuracy: CO:±5%
TB: ±0.1ºC
TI: ±0.1ºC
Parameter output: Cardiac output
Hemodynamics calculation
Multi-gas/O2
Method: Infrared Absorption
Gas accuracy: CO2:0 to 15vol%,
±(0.2vol%+2% of reading) 15 to 25vol%, Unspecified N2O:0~100vol%,
±(2vol%+2% of reading) O2:0~100vol%,±(1vol%+2% of reading)
Enf,Iso,Hal:0~8vol%,
±(0.15vol%+5% of reading) 8 to 25vol%, Unspecified Sev:0~10vol%,±(0.15vol%+5% of reading) 10 to 25vol%, Unspecified Des:0~22 vol%,±(0.15vol%+5% of reading) 22 to 25vol%, Unspecified
Data output: Fi and Etvalues
Respiration rate: 0 to 150 ±1breaths/min
Other: Up to 5 simultaneous gas concentration waveforms, Agent mixture detection, MAC value displayed
Standard Configuration: ECG,SpO2,NIBP,TEMP,PR/HR,RESP
Optional Functions: EtCO2,2IBP,C.O., Multi-gas/O2, Recorder,12 lead ECG