Welcome to Sino-Hero(Shenzhen)Bio-Medical Electronics Co.,Ltd
SV2
Catalog Download
Product Introduction

Smart Multi-Parameter Veterinary Monitor


Display

● 12.1" touch LCD
● Resolution:1024x768 pixels

Power Supply
100~240 VAC, 50/60 Hz, Pmax=90VA Fuse T1.5A

Battery
● 2200mAh 14.8V rechargeable battery
● Operating time after full charge is more than 5 hours
● Operating time after the first alarm of low battery will be about 5 minutes
● Maximum charging time is less than 6 hours.

Signal Interface
● Network interface: Standard RJ45 Socket

Storage
● Trend: 720 hours
● NIBP review: 1000  NIBP events
● Wave review: 2 hours
● Alarm review: 200 alarm events
● All storage data are non volatile.

Environment
● Temperature
Working: 0 ~ 40 ℃
Storage: -20 ~ 50 ℃
● Humidity
Working: 15% ~ 90 %
Storage: 15% ~ 90 % (no coagulation)

ECG
● Lead mode
5 Leads: RA、LA、LL、RL、V; lead mode: I, II, III, AVR, AVL, AVF, V
● Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
● Heart rate
Measure range: 15~300bpm
Accuracy: 1%
Resolution: 1 bpm
● Sensitivity: > 200 μV P-P
● Differential Input Impedance: > 5 M ohm
● Bandwidth
Surgery: 1 ~ 20 Hz
Monitor: 0.5 ~ 40 Hz
Diagnostic: 0.05 ~ 130 Hz
● CMRR
Diagnostic Mode: >90 dB
Monitor Mode: >110 dB
Surgery Mode: >110 dB
● Electrode offset potential: 300mV
● PACE pulse detect
Range: ±4~±700mV
Width: 0.1~2ms
Rise time: 10~100µs
● PACE pulse rejection
Range: ±2~±700mV
Width: 0.1~2ms
Rise time: 10~100µs
● Baseline Recovery: <3 s After defibrillation.
● Signal Range:  8mV p-p
● Calibration Signal: 1mV p-p,   5% accuracy

Respiration
● Method: Impedance between RA-LL
● Respiration Impedance Range: 0.3~3Ω
● Base Impedance Range: 200Ω~4000Ω
● Bandwidth: 0.3 ~ 2.5 Hz
● Gain: x0.25,x0.500,x1,x2,x4
● Respiration Rate
Measurement Range: 0~150BrPM
● Resolution: 1 BrPM
● Accuracy: 0~6 BrPM: unspecified
7~150 BrPM: ±2 BrPM
● Apnea Alarm: 10 ~ 40 s

NIBP
● Method: Oscillometry
● Measure mode: Manual, Auto, STAT
● Measure Interval in AUTO Mode
1,2,3,4,5,10,15,30,60,90,120,180,240,480 min
● Measure Period in STAT Mode: 5 min
● Pulse Rate Range: 40 ~ 240 bpm
● Static pressure accuracy: 3mmHg
● Resolution: 1mmHg
● Accuracy
Maximum Mean error: 5mmHg
Maximum Standard deviation: 8mmHg
● Overpressure Protection: 150 mmHg

SpO2
● Measurement Range: 0 ~ 100 %
● Resolution: 1 %
● Accuracy
70% ~ 100%: ±2 %
<69%: unspecified
● Pulse Rate
Measure and Alarm Range: 20~250bpm
Resolution: 1bpm
Accuracy: 3bpm
● Temperature
Channel: 2
Measure and Alarm Range: 0 ~ 50℃
Resolution: 0.1℃
Accuracy(no sensor): ± 0.1℃(0~ 50℃)
● Recorder(optional)
Paper width: 48 mm
Speed: 25/50 mm/s
Wave channel: 2 channels

Size and Weight
● Product size:305mm×290mm×145mm
● Product weight: 2.80kg
● Carton size: 430x360x280mm
● Carton weight: 4.5kg

Configuration
● 3/5 lead ECG, NIBP, SpO2, RESP, PR, TEMP

Optional
● EtCO2, IBP, CMS, Thermal Printer
Copyright 2017 Sino-Hero(Shenzhen)Bio-Medical Electronics Co.,Ltd. 粤ICP备06126166号